Hach pHD sc Online Process pH Sensor - pH Sensor for Clean Water

Hach pHD sc Online Process pH Sensor -  pH Sensor for Clean Water
Produto n.º: DPD1R1.99
Preço Unitário Contacte-nos
Enviado em 2 semanas

Exceptional Process pH Sensor Performance with the Differential Electrode pHD Measurement Technique

This field-proven technique uses three electrodes instead of the two normally used in conventional pH sensors. Process and reference electrodes measure the pH differentially with respect to a third ground electrode. The end result is unsurpassed measurement accuracy, reduced reference junction potential, and elimination of sensor ground loops. These sensors provide greater reliability, resulting in less downtime and maintenance.

Lower Maintenance Needs with the Double Junction Salt Bridge

The double junction salt bridge creates a barrier to contamination which minimizes the dilution of the internal standard cell solution. The result is lower maintenance needs and a longer time period between calibrations.

Extended Working Life with the Replaceable Salt Bridge/Protector

The unique, replaceable salt bridge holds an extraordinary volume of buffer to extend the working life of the sensor by protecting the reference electrode from harsh process conditions. The salt bridge simply threads onto the end of the sensor if replacement is needed.

Reliability with Built-in Encapsulated Preamp

Encapsulated construction protects the sensor’s built-in preamp from moisture and humidity, ensuring reliable sensor operation. The preamp in the pHD analogue sensor produces a strong signal, enabling the sensor to be located up to 1000 m (3280 ft.) from the analyser.

Innovative Technology

The former GLI, now a Hach Company brand, invented the Differential Electrode Technique for pH measurement in 1970. The pHD takes this field-proven technology to a new level.

Especificações

Compensación de la temperatura: Automatic with NTC 300 Ω thermistor, or manually fixed at a user-entered temperature,
additional selectable temperature correction factors (ammonia, morpholine, or user-defined pH/°C linear slope) available for pure water automatic compensation 0.0 - 50 °C
Compliance: Hazardous location, Maritime, CE
Comprimento: 271.3 mm
Comprimento do cabo: 10 m PUR (polyurethane) 4-conductor with one shield, rated to 105°C
Comunicação: Modbus
Condições de armazenamento: 4 - 70°C, 0-95% relative humidity (non-condensing)
Conexión de cable: Digital
Desvio: 0.03 pH per 24 hours, non-cumulative
Distancia de transmisión: 1000 m (3280 ft.), maximum when used with a termination box.
  máx. 100 m
Exactitud de la temperatura: ± 0,5 °C
Exatidão: ± 0,02 pH
Gama de medição: -2.0 to 14.0 pH
Gama de pressão: Maximum 10.7 bar . 6.9 bar for Digital Sensor at 70°C, and 6.9 bar for Analog Sensor at 105°C.
Garantia: 24 meses
Material da caixa: PPS
Materiales mojados : PEEK or PPS, salt bridge of matching material with PVDF junction, glass process electrode, titanium ground electrode, and FKM/FPM O-ring seals (pH sensor with optional HF-resistant glass process electrode has 316 stainless steel ground electrode, and perfluoroelastomer wetted O-rings; consult factory for other available wetted O-ring materials)
Método de Calibração: Two point automatic, one point automatic, two point manual, one point manual.
Peso: 0.316 kg
Rango de Temperatura de operación: -5 - 70 °C (23 - 158 °F) pHD and ORP
0 - 50 °C (32 - 122 °F) SS pHD

Before initial pH calibration, calibrate the temperature measurement when the sensor is in water or buffer which is at approximately the same temperature as the pH buffers (matches current recommendation)
Repetibilidade: ± 0.05 pH
Rosca del sensor: NPT en ambos extremos
Sensibilidad: ± 0,01 pH
Sensor cable: 10 m (33 ft.) polyurethane, 4-conductor cable with one shield, rated to 105°C (221°F)
Sensor de temperatura: NTC 300 Ω thermistor for automatic temperature compensation and analyzer temperature readout
Taxa de fluxo: 3 m (10 ft.) per second, maximum
Tipo de electrodo: General Purpose
Tipo de montagem: Convertible