Alarmes:
1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Calibração:
Automatic, 2-point; frequency freely programmable
Caudal da amostra:
100 - 300 mL/min
Certificações:
CE compliant / UL certified
Ciclo de medição:
7 - 10 minutes incl. sample lysis
Classe de proteção:
Analyser cabinet: IP44 / Panel PC: IP65
Conteúdos:
EZ7300 ATP Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 2 x empty 5L Reagent Container with Fittings (NaOH & HCl Solution) and 3 x Screwcap/Fittings/Reference1/Reference2
Demineralised water:
For rinsing
Dimensões (A x L x P):
690 mm x 465 mm x 330 mm
Dreno:
Atmospheric pressure, vented, min. Ø 32 mm
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Exactitud:
Better than 4% full scale range for standard test solutions
Fonte de alimentação:
100 - 240 VAC, 50/60 Hz
Gama de medição:
0.5 - 200 pg/mL
0.5 - 200 pg/mL
Garantia:
24 meses
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferencias:
High concentrations of Hg2+, Cu2+, Zn2+, Cd2+, Fe2+. Total salt concentrations higher than 1 g/L. pH lower than 5.5 and higher than 8.
Limite de deteção:
≤ 0.05 pg/mL (0.1 pM) ATP
Material:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Método de medição:
Determination of adenosine triphosphate (ATP) by means of chemiluminescent reaction using luciferin and luciferase, conform with standard method ASTM D4012-81
Modelo:
EZ7300
Número de Corrientes de muestra:
1, 2, 4, or 8
Parâmetro:
ATP
Peso:
30 kg
Potência/Energia:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Pressão da amostra:
Maximum 3 bar for direct injection or by external overflow vessel via the grab sample pump
Reagent Requirements:
Keep between 4 - 8 °C during operation, shielded from light. For prolonged storage, store at -20 °C in a dark environment.
Saídas analógicas:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Saídas digitais:
Optional: Modbus (TCP/IP, RS485)
Temperatura ambiente:
7 - 23 °C ±4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura da amostra:
5 - 30 °C
Validation:
Automatic; frequency freely programmable