Alarmes:
1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Calibração:
N.A.
Caudal da amostra:
100 - 300 mL/min
Certificações:
CE compliant / UL certified
Ciclo de medição:
10 - 15 minutes
Classe de proteção:
Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Conteúdos:
EZ4032 Sulphide Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x Pt Redox Electrode, S7 Connection, 2 x empty 10L Reagent Containers (for Buffer & Na₂S₂O₃ Solution) and 1 x empty 10L Reagent Dark Container with Fittings (for I₂ Solution)
Demineralised water:
For rinsing / dilution
Dimensões (A x L x P):
690 mm x 465 mm x 330 mm
Dreno:
Atmospheric pressure, vented, min. Ø 32 mm
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Exactitud:
Better than 2% full scale range for standard test solutions
Fonte de alimentação:
100 - 240 VAC, 50/60 Hz
Gama de medição:
20 - 1000 mg/L
20 - 1000 mg/L S2-
Optional:
2 - 100 mg/L
5 - 250 mg/L
10 - 500 mg/L
Internal dispenser dilution (max. factor 100)
Garantia:
24 meses
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferencias:
Oxidised forms of manganese interfere. Strong oxidising agents such as chlorine, bromine, chlorine dioxide, iodine, permanganate, hydrogen peroxide and ozone. However reduced forms of these components – bromide, chloride, iodide, manganous ion and oxygen – do not interfere. Reducing agents such as organic sulphites also interfere. Fats, oil, proteins, surfactants and tar.
Limite de deteção:
≤ 2 mg/L
Material:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Método de medição:
Precipitation titration with iodine, conform with standard method APHA 4500-S2 (F)
Número de Corrientes de muestra:
1, 2, 4 or 8
Parâmetro:
Sulphide
Peso:
25 kg
Potência/Energia:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Pressão da amostra:
By external overflow vessel
Qualidade da amostra:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Reagent Requirements:
Keep between 10 - 30 °C
Saídas analógicas:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Saídas digitais:
Optional: Modbus (TCP/IP, RS485)
Temperatura ambiente:
10 - 30 °C ±4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura da amostra:
10 - 30 °C
Validation:
Automatic; frequency freely programmable