EZ4006 Chloride Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x mV electrode, 2 x empty 10L Reagent Containers with Fittings (for Cleaning & Buffer Solution) and 1 x empty 10L Reagent Dark Container with Fittings (for AgNO₃ Solution)

Especificações

Alarmes: 1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Calibração: N.A.
Caudal da amostra: 100 - 300 mL/min
Certificações: CE compliant / UL certified
Ciclo de medição: 10 - 15 minutes
Classe de proteção: Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Conteúdos: EZ4006 Chloride Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x mV electrode, 2 x empty 10L Reagent Containers with Fittings (for Cleaning & Buffer Solution) and 1 x empty 10L Reagent Dark Container with Fittings (for AgNO₃ Solution)
Demineralised water: For rinsing / dilution
Dimensões (A x L x P): 690 mm x 465 mm x 330 mm
Dreno: Atmospheric pressure, vented, min. Ø 32 mm
Earth connection: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Exactitud: Better than 2% full scale range for standard test solutions
Fonte de alimentação: 100 - 240 VAC, 50/60 Hz
Gama de medição: 25 ‐ 500 mg/L
  25 - 500 mg/L Cl-

Optional:
5 - 50 mg/L Cl-
5 - 125 mg/L Cl-
5 - 250 mg/L Cl-
Internal dispenser dilution (max. factor 100)
Garantia: 24 meses
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferencias: Bromide, sulphide, iodide ions may interfere. Ferricyanide causes high results and must be removed. Chromate and dichromate interfere and should be reduced to chromic state or removed. Ferric iron interferes if present in an amount substantially higher than the amount of chloride. Chromic ion, ferrous ion and phosphate do not interfere. Fats, oil, proteins, surfactants and tar.
Limite de deteção: 5 mg/L
Material: Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Metodo: Titrimetric
Método de medição: Potentiometric titration with silver nitrate (AgNO₃), conform with standard method APHA 4500-Cl (D)
Modelo: EZ4006
Número de Corrientes de muestra: 1, 2, 4 or 8
Parâmetro: Chloride
Peso: 25 kg
Potência/Energia: 100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Pressão da amostra: By external overflow vessel
Qualidade da amostra: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Reagent Requirements: Keep between 10 - 30 °C
Saídas analógicas: Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Saídas digitais: Optional: Modbus (TCP/IP, RS485)
Temperatura ambiente: 10 - 30 °C ±4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura da amostra: 10 - 30 °C
Validation: Automatic; frequency freely programmable