EZ3505sc Chloride Ion-Selective Electrode (ISE) Standard Addition Analyser, 1 stream, 4x mA

EZ3505sc Chloride Ion-Selective Electrode (ISE) Standard Addition Analyser, 1 stream, 4x mA
Produto n.º: EZ3505.9800140T
Preço Unitário Contacte-nos

Advanced Standard Addition Technology

Ensure reliable data accuracy with advanced discontinuous and standard addition modes. This feature automatically compensates for interferences, delivering precise results even in complex or changing water conditions. Avoid compliance risks caused by inconsistent or unreliable measurements.

Designed for Low-Maintenance Operation

Frequent manual calibration and sensor cleaning increase maintenance effort and downtime. The EZ3505sc employs discontinuous and automated standard addition modes to compensate for matrix effects, stabilise potentiometric response, reduce reagent consumption, and maximise analyser uptime for reliable chloride monitoring.

Seamless System Integration

Isolated data silos hinder effective plant management. Featuring standard analogue and digital communication protocols, this analyser integrates easily into existing plant networks. Real-time visibility ensures you can make informed decisions quickly based on current process conditions.

Cost-Effective Scalability

Monitoring multiple sample points traditionally requires significant capital investment. This system supports multi-stream functionality, allowing a single unit to analyse up to 8 sample streams. You gain comprehensive process insight while optimising infrastructure costs and reducing equipment footprint.

Conteúdos

EZ3505sc Chloride Ion-Selective Electrode (ISE) Standard Addition Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Electrode and Chloride Electrode, combined, and 2 x empty 5L Reagent Containers with Fittings (for Buffer and Standard Cl Solution)

Especificações

Calibração: Automatic; 2-point; offset or slope; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced
Caudal da amostra: 100 - 300 mL/min
Certificações: CE, ETL certified to UL and CSA safety standards, UKCA
Ciclo de medição: No dilution
Default: 5 minutes
Continuous: 4 minutes

Internal dilution
Default: 10 minutes
Continuous: 7 minutes
Classe de proteção: IP44
Conteúdos: EZ3505sc Chloride Ion-Selective Electrode (ISE) Standard Addition Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Electrode and Chloride Electrode, combined, and 2 x empty 5L Reagent Containers with Fittings (for Buffer and Standard Cl Solution)
Demineralised water: Internal dilution option: Rinse water and Dilution water
Dimensões (A x L x P): 688 mm x 460 mm x 340 mm
Dreno: Atmospheric pressure, vented, min. Ø 32 mm
Earth connection: Earth connection Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Exactitud: Better than 2% full scale range for standard test solutions
Gama de medição: No dilution
2 - 100 mg/L Clˉ
5 - 250 mg/L Clˉ
10 - 500 mg/L Clˉ
20 - 1000 mg/L Clˉ

Internal dilution
100 - 5000 mg/L Clˉ
200 - 10000 mg/L Clˉ
Garantia: 24 meses
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar
Interferencias: Bromide [(Br)-], sulphide [(S)²-], iodide [(I)-], cyanide [(CN)-] ions may interfere. Mercury [(Hg)+] must be absent. Ammonia [NH₃] and thiosulphate [(S₂O₃)²-] may interfere. Fats, oil, proteins, surfactants and tar.
Limite de deteção: No dilution
2 - 100 mg/L Clˉ: 2 mg/L
5 - 250 mg/L Clˉ: 5 mg/L
10 - 500 mg/L Clˉ: 10 mg/L
20 - 1000 mg/L Clˉ: 20 mg/L

Internal dilution
100 - 5000 mg/L Clˉ: 100 mg/L
200 - 10000 mg/L Clˉ: 200 mg/L
Limpeza automática: Yes; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
Material: Hinged part: Thermoform ABS, Door: PMMA
Wall section: Galvanised steel, powder coated
Método de medição: Discontinuous, direct measurement by combined Ion-Selective Electrode with standard addition
  Ion-Selective Electrode (standard addition)
Número de Corrientes de muestra: Standard: 1
Optional accessory for 2, 4, or 8 sample streams
Parâmetro: Chloride
Peso: Max. 35 kg
Potência/Energia: 100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Pressão da amostra: By external overflow vessel (open to atmospheric pressure)
Qualidade da amostra: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Reagent Requirements: Keep between 10 - 30 °C
Saídas analógicas: Active 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard 4, optional: 8
Saídas digitais: Relays: 5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150)

Ethernet Connections: Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP

RS485 communication: Profibus DP or Modbus RTU
Temperatura ambiente: 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura da amostra: 10 - 30 °C
Validation: Automatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly