Alarmes:
1x malfunctioning, 4x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Calibração:
Automatic, 2-point; frequency freely programmable
Caudal da amostra:
100 - 300 mL/min
Certificações:
CE compliant / UL certified
Ciclo de medição:
8 minutes
Classe de proteção:
Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Conteúdos:
EZ3503 Chloride Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Electrode and Chloride Electrode, combined, and 2 x empty 5L Reagent Containers with Fittings (for Buffer and Standard Cl Solution)
Demineralised water:
For rinsing
Dimensões (A x L x P):
690 mm x 465 mm x 330 mm
Dreno:
Atmospheric pressure, vented, min. Ø 32 mm
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Exactitud:
Better than 2% full scale range for standard test solutions
Fonte de alimentação:
100 - 240 VAC, 50/60 Hz
Gama de medição:
1 ‐ 10 mg/L
Garantia:
24 meses
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferencias:
Bromide, sulphide, iodide, cyanide ions may interfere. Mercury must be absent. Ammonia and thiosulphate may interfere. Fats, oil, proteins, surfactants and tar.
Limite de deteção:
1 mg/L
Limpeza automática:
Sim
Material:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Metodo:
Ionselective (standard addition)
Método de medição:
Discontinuous measurement by combined ion-selective electrode with standard addition
Modelo:
EZ3503
Número de Corrientes de muestra:
1, 2, 4 or 8
Parâmetro:
Chloride
Peso:
25 kg
Potência/Energia:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Pressão da amostra:
By external overflow vessel
Qualidade da amostra:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Reagent Requirements:
Keep between 10 - 30 °C
Saídas analógicas:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Saídas digitais:
Optional: Modbus (TCP/IP, RS485)
Temperatura ambiente:
10 - 30 °C ±4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura da amostra:
10 - 30 °C
Validation:
Automatic; frequency freely programmable