EZ3005sc Chloride Ion-Selective Electrode (ISE) Analyser, 1 stream, 4x mA

EZ3005sc Chloride Ion-Selective Electrode (ISE) Analyser, 1 stream, 4x mA
Produto n.º: EZ3005.9800140T
Preço Unitário Contacte-nos

Reliable, Stable Measurements

Avoid downtime from unreliable ISE data caused by temperature fluctuations. The EZ3005sc stabilises both sample and electrode using a temperature controlled flow cell, eliminating measurement errors and delivering precise, repeatable chloride results operators can trust.

Lower Maintenance Effort

Minimise downtime and service costs with the EZ3005sc. Automated 2 point calibration, reagent alerts, a rugged ISE electrode, and minimal moving parts extend maintenance intervals; so operators spend less time servicing instruments and more time on plant operations.

Reduced Operating Cost

The EZ3005sc helps optimise operating costs by reducing chemical consumption and service visits. Multi stream capability monitors up to eight sample lines, lowering cost per sampling point and extending equipment life.

Designed for easy integration into existing control systems

With a wide range of analogue and digital outputs, this analyser aligns with plant specific requirements, ensuring compatibility with existing infrastructure and immediate access to real time measurement data.

Conteúdos

EZ3005sc Chloride Ion-Selective Electrode (ISE) Analyser, 1 stream, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Electrode and Chloride Electrode, combined, and 1 x empty 2.5L Reagent Container with Fittings (for Buffer Solution)

Especificações

Calibração: Automatic; 2-point; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced
Caudal da amostra: 100 - 300 mL/min
Certificações: CE, ETL certified to UL and CSA safety standards, UKCA
Ciclo de medição: Default: 5 minutes
Continuous: 5 minutes
Classe de proteção: IP44
Conteúdos: EZ3005sc Chloride Ion-Selective Electrode (ISE) Analyser, 1 stream, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Electrode and Chloride Electrode, combined, and 1 x empty 2.5L Reagent Container with Fittings (for Buffer Solution)
Demineralised water: Does not apply
Dimensões (A x L x P): 688 mm x 460 mm x 340 mm
Dreno: Atmospheric pressure, vented, min. Ø 32 mm
Earth connection: Earth connection Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Exactitud: Better than 2% full scale range for standard test solutions
Gama de medição: 2 - 100 mg/L Clˉ
5 - 250 mg/L Clˉ
10 - 500 mg/L Clˉ
20 - 1000 mg/L Clˉ
Garantia: 24 meses
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar
Interferencias: Bromide [(Br)-], sulphide [(S)²-], iodide [(I)-], cyanide [(CN)-] ions may interfere. Mercury [(Hg)+] must be absent. Ammonia [NH₃] and thiosulphate [(S₂O₃)²-] may interfere. Fats, oil, proteins, surfactants and tar.
Limite de deteção: 2 - 100 mg/L Clˉ: 2 mg/L
5 - 250 mg/L Clˉ: 5 mg/L
10 - 500 mg/L Clˉ: 10 mg/L
20 - 1000 mg/L Clˉ: 20 mg/L
Limpeza automática: Yes; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
Material: Hinged part: Thermoform ABS, Door: PMMA
Wall section: Galvanised steel, powder coated
Método de medição: Discontinuous, direct measurement by combined Ion-Selective Electrode
  Ion-Selective Electrode
Número de Corrientes de muestra: Standard: 1
Optional accessory for 2, 4, or 8 sample streams
Parâmetro: Chloride
Peso: Max. 35 kg
Potência/Energia: 100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Pressão da amostra: By external overflow vessel (open to atmospheric pressure)
Qualidade da amostra: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Reagent Requirements: Keep between 10 - 30 °C
Saídas analógicas: Active 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard 4, optional: 8
Saídas digitais: Relays: 5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150)

Ethernet Connections: Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP

RS485 communication: Profibus DP or Modbus RTU
Temperatura ambiente: 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura da amostra: 10 - 30 °C
Validation: Automatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly