EZ2300 Total Aluminium + Aluminium Al(III) Analyser

EZ2300 Total Aluminium + Aluminium Al(III) Analyser
Produto n.º: EZ2300.XXXXXXXX
Preço Unitário Contacte-nos

Conteúdos

EZ2300 Aluminum Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets and 4 x empty 2.5L Reagent Containers with Fittings (for Acid, Buffer, Colour & Mixed Reagent Solution)

Especificações

Alarmes: 1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Calibração: Automatic, 2-point; frequency freely programmable
Caudal da amostra: 100 - 300 mL/min
Certificações: CE compliant / UL certified
Ciclo de medição: 20 min Total Al (dilution + 5 min)
30 min Total Al & Al (III)
Classe de proteção: Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Conteúdos: EZ2300 Aluminum Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets and 4 x empty 2.5L Reagent Containers with Fittings (for Acid, Buffer, Colour & Mixed Reagent Solution)
Demineralised water: For rinsing / dilution
Dimensões (A x L x P): 690 mm x 465 mm x 330 mm
Dreno: Atmospheric pressure, vented, min. Ø 32 mm
Earth connection: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Exactitud: Better than 2% full scale range for standard test solutions
Gama de medição: 0 - 150 µg/L Al
  10 - 150 µg/L Al

Optional:
10 - 75 µg/L
80 - 600 µg/L (with internal dilution)
160 - 1500 µg/L (with internal dilution)
Garantia: 24 meses
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferencias: Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Limite de deteção: ≤ 10 µg/L
Material: Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Método de medição: Colorimetric measurement using pyrocatechol violet method at 578 nm
Modelo: EZ2300
Número de Corrientes de muestra: 1, 2, 4 or 8
Parâmetro: Aluminum, total; Al(III), dissolved
Peso: 25 kg
Potência/Energia: 230 VAC, 50/60 Hz
120 VAC, 50/60 Hz
Max. power consumption: 440 VA
Pressão da amostra: By external overflow vessel
Reagent Requirements: Keep between 10 - 30 °C
Saídas analógicas: Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Saídas digitais: Optional: Modbus (TCP/IP, RS485)
Temperatura ambiente: 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura da amostra: 10 - 30 °C
Validation: Automatic; frequency freely programmable