EZ1024 Iron Analyser Fe(II+III)

EZ1024 Iron Analyser Fe(II+III)
Produto n.º: EZ1024.XXXXXXXX
Preço Unitário Contacte-nos

Conteúdos

EZ1024 Iron Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 2 x empty 5L Reagent Container with Fittings (for Buffer & Reductor Solution) and 1 x empty 2.5L Glass Reagent Container with Fittings (for Colour Solution)

Especificações

Alarmes: 1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Calibração: Automatic, 2-point; frequency freely programmable
Caudal da amostra: 100 - 300 mL/min
Certificações: CE compliant / ETL certified
Ciclo de medição: 10 min Fe (II), Fe total dissolved (dilution + 5 min.)
15 min all combined parameters
Classe de proteção: Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Conteúdos: EZ1024 Iron Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 2 x empty 5L Reagent Container with Fittings (for Buffer & Reductor Solution) and 1 x empty 2.5L Glass Reagent Container with Fittings (for Colour Solution)
Demineralised water: For rinsing / dilution
Dimensões (A x L x P): 690 mm x 465 mm x 330 mm
Dreno: Atmospheric pressure, vented, min. Ø 32 mm
Earth connection: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Exactitud: Better than 2% full scale range for standard test solutions
Fonte de alimentação: 100 - 240 VAC, 50/60 Hz
Gama de medição: 0.01 - 1 mg/L Fe(II)
  0.01 - 1 mg/L Fe(II)

Optional:
0.005 - 0.25 mg/L
0.005 - 0.5 mg/L
0.08 - 4 mg/L (with internal dilution)
0.16 - 8 mg/L (with internal dilution)
2 - 100 mg/L (with internal dilution)
Garantia: 24 meses
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferencias: Metal ions like Lead > 10 mg/L, Zinc > 2 mg/L, Nickel > 2 mg/L, Iron > 5 mg/L, Copper > 5 mg/L. Strong oxidising agents, Cyanide, Nitrite, Phosphate (polyphosphate more than orthophosphate), Chromium, Zinc in concentrations exceeding 10 times that of Iron. Bismuth, Cadmium, Mercury, Molybdate, and Silver precipitate Phenanthroline. Polyphosphate must be absent. Large amounts of colour and turbidity interfere. Fats, pil, proteins,surfactants and tar.
Limite de deteção: ≤ 2 µg/L
Limpeza automática: Sim
Material: Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Método de medição: Colorimetric measurement using TPTZ colour solution, conform with APHA 3500-Fe (B)
Modelo: EZ1024
Número de Corrientes de muestra: 1, 2, 4, or 8
Parâmetro: Iron Fe(II+III), total dissolved
Peso: 25 kg
Potência/Energia: 100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Pressão da amostra: By external overflow vessel
Qualidade da amostra: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Reagent Requirements: Keep between 10 - 30 °C
Saídas analógicas: Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Saídas digitais: Optional: Modbus (TCP/IP, RS485)
Temperatura ambiente: 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura da amostra: 5 - 30 °C
Validation: Automatic; frequency freely programmable