Alarmes:
1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Calibração:
Automatic, 2-point; frequency freely programmable
Caudal da amostra:
100 - 300 mL/min
Certificações:
CE compliant / ETL certified
Ciclo de medição:
10 min (dilution + 5 min.)
Classe de proteção:
Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Conteúdos:
EZ1022 Hydrogn Peroxide Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets and 2 x empty 2.5L Glass Containers with Fittings (for Colour and Buffer Solution)
Demineralised water:
For rinsing / dilution
Dimensões (A x L x P):
690 mm x 465 mm x 330 mm
Dreno:
Atmospheric pressure, vented, min. Ø 32 mm
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Exactitud:
Better than 2% full scale range for standard test solutions
Fonte de alimentação:
100 - 240 VAC, 50/60 Hz
Gama de medição:
5 - 500 µg/L H₂O₂
5 - 500 µg/L H₂O₂
Optional:
2 - 125 µg/L
2 - 250 µg/L
0.04 - 2.0 mg/L (with internal dilution)
0.08 - 4.0 mg/L (with internal dilution)
0.5 - 50 mg/L (with internal dilution)
Garantia:
24 meses
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferencias:
Oxidants like Bromine, Bromamine, Chlorine, Copper Cu(II) > 8 mg/L, Chromate, Iodine, Iodoamines, Iron Fe(III) > 20 mg/L, Nitrite, Ozone and Manganese may oxidize the DPD colour solution. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Limite de deteção:
≤ 2 µg/L
Limpeza automática:
Sim
Material:
Hinged part: Thermoform ABS, door: PMMA
Wall section: galvanised steel, powder coated
Método de medição:
Colorimetric measurement at 510 nm using DPD
Modelo:
EZ1022
Número de Corrientes de muestra:
1, 2, 4, or 8
Parâmetro:
Hydrogen Peroxide
Peso:
25 kg
Potência/Energia:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Pressão da amostra:
By external overflow vessel
Qualidade da amostra:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Reagent Requirements:
Keep refrigerated.
Saídas analógicas:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Saídas digitais:
Optional: Modbus (TCP/IP, RS485)
Temperatura ambiente:
10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura da amostra:
10 - 30 °C
Validation:
Automatic; frequency freely programmable