Alarmes:
                                            
                                            
                                        
									    1x malfunctioning, 4x user-configurable, max. 24 VDC/0.5 A, potential free contacts
                                    
                                
    						
								
                                    
                                        
                                            
                                                Calibração:
                                            
                                            
                                        
									    Automatic, 2-point; frequency freely programmable
                                    
                                
    						
								
                                    
                                        
                                            
                                                Caudal da amostra:
                                            
                                            
                                        
									    100 - 300 mL/min
                                    
                                
    						
								
                                    
                                        
                                            
                                                Certificações:
                                            
                                            
                                        
									    CE compliant / ETL certified
                                    
                                
    						
								
                                    
                                        
                                            
                                                Ciclo de medição:
                                            
                                            
                                        
									    10 min (dilution + 5 min)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Classe de proteção:
                                            
                                            
                                        
									    Protection class: Analyser cabinet: IP44 / Panel PC: IP65
                                    
                                
    						
								
                                    
                                        
                                            
                                                Conteúdos:
                                            
                                            
                                        
									    EZ1005 Chloride Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 2 x empty 2.5L Reagent Container with Fittings (for Acid & Cleaning Solution) and 1 x empty 2.5L Glass Reagent Container with Fittings (for AgNO₃ Solution)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Demineralised water:
                                            
                                            
                                        
									    For rinsing / dilution
                                    
                                
    						
								
                                    
                                        
                                            
                                                Dimensões (A x L x P):
                                            
                                            
                                        
									    690 mm x 465 mm x 330 mm
                                    
                                
    						
								
                                    
                                        
                                            
                                                Dreno:
                                            
                                            
                                        
									    Atmospheric pressure, vented, min. Ø 32 mm
                                    
                                
    						
								
                                    
                                        
                                            
                                                Earth connection:
                                            
                                            
                                        
									    Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
                                    
                                
    						
								
                                    
                                        
                                            
                                                Exactitud:
                                            
                                            
                                        
									    Better than 2% full scale range for standard test solutions
                                    
                                
    						
								
                                    
                                        
                                            
                                                Fonte de alimentação:
                                            
                                            
                                        
									    100 - 240 VAC, 50/60 Hz
                                    
                                
    						
								
                                    
                                        
                                            
                                                Gama de medição:
                                            
                                            
                                        
									    1 ‐ 10 mg/L
                                    
                                
                                    
                                        
                                            
                                            
                                                 
                                            
                                        
									    1 - 10 mg/L Cl-
Optional:
4 - 40 mg/L Cl- (with internal dilution)
8 - 80 mg/L Cl- (with internal dilution)
100 - 1000 mg/L Cl- (with internal dilution)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Garantia:
                                            
                                            
                                        
									    24 meses
                                    
                                
    						
								
                                    
                                        
                                            
                                                Instrument air:
                                            
                                            
                                        
									    Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
                                    
                                
    						
								
                                    
                                        
                                            
                                                Interferencias:
                                            
                                            
                                        
									    Substances in amounts normally found in potable water will likely not interfere. Bromide, iodidie and cyanide register as equivalent chloride concentrations. Sulfide, thiosulfate and sulfite interfere but can be removed by treatment with hydrogen peroxide. Orthophosphate > 25 mg/L interferes by precipitating as silver phosphate. Iron > 10 mg/L interferes. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
                                    
                                
    						
								
                                    
                                        
                                            
                                                Limite de deteção:
                                            
                                            
                                        
									    1 mg/L
                                    
                                
    						
								
                                    
                                        
                                            
                                                Limpeza automática:
                                            
                                            
                                        
									    Sim
                                    
                                
    						
								
                                    
                                        
                                            
                                                Material:
                                            
                                            
                                        
									    Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
                                    
                                
    						
								
                                    
                                        
                                            
                                                Método de medição:
                                            
                                            
                                        
									    Colorimetric measurement at 546 nm using diphenylcarbazide method, conform with standard method APHA 3500-Cr (B)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Modelo:
                                            
                                            
                                        
									    EZ1005
                                    
                                
    						
								
                                    
                                        
                                            
                                                Nome do método:
                                            
                                            
                                        
									    Colorimetric
                                    
                                
    						
								
                                    
                                        
                                            
                                                Número de Corrientes de muestra:
                                            
                                            
                                        
									    1, 2, 4, or 8
                                    
                                
    						
								
                                    
                                        
                                            
                                                Parâmetro:
                                            
                                            
                                        
									    Chloride
                                    
                                
    						
								
                                    
                                        
                                            
                                                Peso:
                                            
                                            
                                        
									    25 kg
                                    
                                
    						
								
                                    
                                        
                                            
                                                Potência/Energia:
                                            
                                            
                                        
									    100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
                                    
                                
    						
								
                                    
                                        
                                            
                                                Pressão da amostra:
                                            
                                            
                                        
									    By external overflow vessel
                                    
                                
    						
								
                                    
                                        
                                            
                                                Qualidade da amostra:
                                            
                                            
                                        
									    Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
                                    
                                
    						
								
                                    
                                        
                                            
                                                Reagent Requirements:
                                            
                                            
                                        
									    Keep between 10 - 30 °C
                                    
                                
    						
								
                                    
                                        
                                            
                                                Saídas analógicas:
                                            
                                            
                                        
									    Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Saídas digitais:
                                            
                                            
                                        
									    Optional: Modbus (TCP/IP, RS485)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Temperatura ambiente:
                                            
                                            
                                        
									    10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Temperatura da amostra:
                                            
                                            
                                        
									    10 - 30 °C
                                    
                                
    						
								
                                    
                                        
                                            
                                                Validation:
                                            
                                            
                                        
									    Automatic; frequency freely programmable