510 three channels

510 three channels
Produto n.º: 510CCC/W2C0P000
Preço Unitário Contacte-nos
Enviado em 2 semanas

Easy to use, intuitive software using clear, full color touch-screen

All functions of Orbisphere 510 instruments are accessed through the color, touch screen.This screen acts as display and keyboard.

Internal diagnostics and reminders for maintenance and calibration

To ensure continuous high performance and simplify maintenance, Orbisphere 510 instruments offer a number of diagnostic features.
Notification that calibration is due—ensures QC procedural compliance
Notification that a sensor service is due—optimal preventative maintenance planning
Sensor service diagnostics—minimizes downtime
System alarms transmitted through analog output—continuous status indication

Easy to use and reliable

Orbisphere 510 instruments are designed for ease of use. After simple installation, the system is measuring your process in minutes. Simple menus allow you to rapidly create a specific configuration for the instrument. The main measurement window continuously displays real time process readings, graphed sensor trends (user selectable from last 1 minute to last 1 hour), alarm limits, temperature and event occurrence.

Multiple communication options including USB and Profibus

Digital communication uses industry standard protocols including Profibus and USB in addition to traditional analog outputs and alarm relays. These may all be configured in terms of function, content and behavior.

Data Storage

Data storage of up to 10000 measurements, last 1000 operator actions and details of last 50 calibrations

Conteúdos

510 controller and manual only.
Sensors and cables, flow-chamber or process connection, must be ordered separately.

Especificações

Capacidades de comunicação: 3 x 0/4-20 mA; RS485; Ethernet
Certificação de conformidade: EMC: EN61326-1:2006
CE: EN61010-1:2010
ETL, conforming to UL 61010-1 and CSA 22.2 No. 61010-1
Comunicação: 3 x 0/4-20 mA; RS485; Ethernet
Condições de operação: 0 - 95 % non-condensing relative humidity
Construção da carcaça: Wall mount
Conteúdos: 510 controller and manual only.
Sensors and cables, flow-chamber or process connection, must be ordered separately.
Controlador: 510, Multi-Channel
Dados armazenados: Rolling buffer or store once mode for up to 1000 measurements and 1000 operator actions

Holds calibration records for last 50 calibrations
Dimensões (A x L x P): 230.5 mm x 250 mm x 160 mm
Ecrã: Colour TFT touchscreen display
Equipamentos compativeís: Channel 1 : 313xx or C1100
Channel 2 : 313xx or C1100
Channel 3 : 313xx or C1100
Exatidão: Accuracy: ± 0.5% (between operating temperature limits)
Fonte de alimentação: 10 to 30 VDC - 30W
Interfaz de usuario: Touch screen panel displays: concentration, trend graph, diagnostics, alarm status, historical data
Material: Stainless steel
Medições: Resolution, accuracy and response time are determined by sensor
Número de canais: 3
Parâmetro: O₃ (EC), O₃ (EC), O₃ (EC)
Peso: 3.8 kg
Potência/Energia: 10 to 30 VDC - 30W
Proteção por senha: 5 levels of authorised access to configuration and data management
Relés: 3 measurement alarm relays (2A to 30 VAC or 0.5 A to 50 VDC); configurable to send diagnostics information
1 system alarm relay (2 A to 30 VAC or 0.5 A to 50 VDC)
Requisitos de alimentação (tensão): 10 - 30 V DC
Saídas analógicas: 3 smart 0/4-20 mA (500 ohms), programmable as linear or tri-liniear, configurable to send diagnostics information
Temperatura de funcionamento: -5 - 40 °C
Unidades: • Gas concentration: Configurable for gas or liquid phase with multiple unit options
• Pressure : External and barometric pressure with multiple unit options
• Temperature : Sample temperature with unit options (K, °C, °F)