Polymetron 2586sc Oxygen Scavenger Analyser, Smart Monitoring, 5x mA Output, 110-240 VAC, without power cord

Polymetron 2586sc Oxygen Scavenger Analyser, Smart Monitoring, 5x mA Output, 110-240 VAC, without power cord
Produto n.º: 2586.99.AA1801
Preço Unitário Contacte-nos
Enviado em 2 semanas

Save Time on Design

A single design source and one product platform means you spend less time searching for design files or configuring components.

Easy Adoption

The familiar experience with a modern touchscreen, the ability to use your current Hach sensor, and the same footprint as the 9500 Polymetron, make installation and integration of the SC4500 Controller seamless.

The Connectivity Options You Need

The Controller provides local communication to SCADA or a PLC. From analog, to advanced digital protocols, to wi-fi, cellular or LAN, the SC4500 gives you the flexibility to adapt in a rapidly changing world.

Simplify Maintenance and Operation

Common menu guides reduce variability and provide step-by-step procedures for maintenance and calibration. Standard visual alerts across parameters notify operators when troubleshooting is required. The Hach Polymetron 2586sc has a fast response time of less than 60 seconds.
Unlike traditional amperometric techniques that use two electrodes, the Hach Polymetron 2586sc oxygen scavenger analyser uses a three-electrode design; eliminating voltage drift due to the composition of the water. Self-cleaning electrodes reduce maintenance costs and analyser downtime via PTFE beads that prevent deposits on the electrode surfaces.

Conteúdos

Controller installed in Panel, Working Electrode, Reference Electrode, Counter Electrode, Basic User Manual

Especificações

Capacidades de comunicação: Claros mA Output
Certificaciones eléctricas: EMC

CE compliant for conducted and radiated emissions:
- CISPR 11 (Class A limits)
- EMC Immunity EN 61326-1 (Industrial limits)

Safety
CAN/CSA C22.2 No. 61010-1

cETLus safety mark for:
- General Locations per ANSI/UL 61010-1 & CAN/CSA C22.2. No. 61010-1
Comunicación (opcional): Analog:
Five 0-20 mA or 4-20 mA analog outputs on each analog output module
Up to two analog Input modules (0-20 mA or 4-20 mA). Each input module replaces a digital sensor input.

Digital:
Profibus DPV1 module
Modbus TCP
Profinet IO module
Ethernet IP module
Condições de armazenamento: -20 - 70 °C, 0 - 95% relative humidity, non-condensing
Conexión: 4 x 6 mm stainless steel tubing
  6 x 8 mm (tubing must not exceed 4 feet and must drain straight down)
Conteúdos: Controller installed in Panel, Working Electrode, Reference Electrode, Counter Electrode, Basic User Manual
Fluxo: 166 - 250 mL/min (10 - 15 L/h) recommended
Fonte de alimentação: Without power cord
Funções de relé: Scheduler (Timer), Alarm, Feeder Control, Event Control, Pulse Width Modulation, Frequency Control, and Warning
Gama de medição: 0 - 500 ppb hydrazine; programmable
  0 - 500 ppb hydrazine; programmable
Gama de medição (2): 0 - 100 ppb carbohydrazide (also known as ELIMIN-OX); programmable

ELIMIN-OX is a registered trademark of Nalco Chemical Co., Naperville, IL.
Gama de pressão: 0.5 - 6 bar (7.2 - 87 psi) or 12 L/h
Humedad relativa: 10 - 90% relative humidity (non-condensing)
Idiomas del manual: Bulgarian, Chinese (PRC), Croatian, Czech, Danish, Dutch, English, Estonian, Finnish, French, German, Greek, Hungarian, Italian, Japanese, Korean, Lithuanian, Polish, Portuguese (Brazil), Portuguese (Portugal), Romanian, Russian, Slovak, Slovenian, Spanish, Swedish, Thai, Turkish
Intervalo de manutenção: Monthly: Calibration and reagent refill
Limite de deteção: Drift is negligible; 1 ppb
Material: Counter-Electrode: Stainless steel; Working Electrode: Platinum
Número de canais: 1
Peso: 14.6 kg
Reactivo: Sample conditioning with diisopropylamine, ammonia, or diethylamine
Referencia: Ag/AgCl electrode with electrolyte KCl 0.1 M
Relé: Modo operacional: Primary or secondary measurement or timer/scheduler
Relés: Four electromechanical SPDT (Form C) contacts, 1200 W, 5 A
Repetibilidade: ± 2% or 1 ppb (whichever is greater)
Requisitos da amostra: Sample needs to be free of undissolved matter.
Requisitos de alimentação (tensão): 110 - 240 V AC
Security levels: 2
Software disponível: Cloud (Claros-enabled)
Temperatura da amostra: 5 - 45 °C
Temperatura de funcionamento: 5 - 45 °C at 0 - 95% relative humidity (non-condensing)
Tempo de resposta: < 60 s